Paper Sealer Test (RRS 14-81110)


Due to the increasing use of disposable paper packaging (W-Pack) in medical centers and the importance of sealing, it is necessary to use a sealer device to seal this type of packaging paper. The sealer device must be evaluated daily. When we put an instrument in disposable papers and sterilize it in an autoclave, we must make sure that after removing the package until it is used in the surgical room, the ambient air does not enter the package and does not take it out of the sterile state. Therefore, the correct operation of the sealing and packaging machine should be evaluated. The Paper Sealer device Test (RRS 14-81110) is a standard and ideal method for diagnosing porosity and correct sealing in paper and sterile medical packages (V-pack).

Instruction for use 

  • Turn on the sealer device and allow it to reach to its appropriate operating temperature
  • Separate about 10 cm from the roll-pack for sterile packaging.
  • Place a sheet of sealer test between the two layers of packages and pass the package pack through the sealer device

Result

The part that passes through the device must have a continuous black line. The continuous line should be darker than the other parts. No bubbles should be seen in this section.

Note

The quality of the seal stages should be validated using ink and paper tests and then archived. Having a regular monitoring program helps to quickly detect any changes or problems in the packaging process.

Advantages

  • Confirm that the packages are sterile during storage
  • Ensure the correct sealing of the sealer device
  • Quick and easy to use
  • Low costs