Helix PCD (RRS 14-11510)


Helix PCD (RRS 14-11510) is used to monitor a set of devices and packages in a steam autoclave, ethylene oxide, and formaldehyde. This test is used to ensure the sterilization process of challenging loads such as long tubes with small diameter, the internal surface of devices, and porous items have been done successfully.

The Helix PCD contains a stainless steel chamber and a helix tube (1.5 meters). The test strip is placed in the chamber. 

Instructions for RRS Helix PCD 

  • The 4-dot test strip should bend in the middle. the part with the indicator is placed inwards (Figure 1).
  • Insert the bent part into the groove of the plastic piece (Figure 2).
  • Close the cap of the metal container (Figure 3).
  • Place the Bowie-Dick containing the indicator in the surgical towels (Figure 4).
Instructions for Helix PCD (RRS 14-11510)

Note: The surgical towels is used only for the steam autoclave.

Place the PCD and the surgical towels in the empty autoclave chamber. When the autoclave cycle is done, take out the 4-dot indicator and evaluate it. The color change is permanent and the test strips can be documented.

How does an autoclave work?

At the beginning of the sterilization period, the vacuum created by the device evacuates the air inside the device and the helical tube. In the second stage, with the entry of sterilizing agents (steam, FO, and EO gases) into the spiral tube and then the chamber, the 4-dots indicator inside the chamber or test pack changes color. At the end of the sterilization cycle, the color of the indicator strip is evaluated.

Result

The yellow color of the indicator has to turn black, fully to make sure of the sterilizer’s accurate function. If the color does not change or brown spots are observed, the device must be controlled in terms of the mentioned parameters.

The Evaluated parameters

  • Air leakage
  • Incompatible gases
  • Insufficient Airflow
  • Vapor penetration into the interior surfaces of equipment such as narrow pipes, pipes, and perforated fittings

Download Catalog

Please Check out file at the following link

Compact Helix PCD (RRS 14-11290)

Ethylene Oxide Category Control indicator (PCD) (RRS 14-12520)

All in one Helix PCD (RRS 14-11280)

Formaldehyde Category Control indicator (PCD) (RRS 14-14520)

Process Challenge Devices (RRS 14-11520) (PCD)

×